Model 9110TH NOx Analyzer

Overview of Operating Modes

4.1.2. WARNING MESSAGES

The most common and serious instrument failures will activate Warning Messages that are displayed on the analyzer’s Front Panel. They are listed on Table 4-3 as follows:

Table 4-3: Warning Messages Defined

MESSAGE

MEANING

ANALOG CAL WARNING

The A/D or at least one D/A channel has not been calibrated.

AZERO WARN

Auto-zero reading above limit specified by AZERO_LIMIT variable. Value shown

 

in message indicates auto-zero reading at time warning was displayed.

BOX TEMP WARNING

The temperature inside the 9110T chassis is outside the specified limits.

CANNOT DYN SPAN

Contact closure span calibration failed while DYN_SPAN was set to ON.

CANNOT DYN ZERO

Contact closure zero calibration failed while DYN_ZERO was set to ON.

CONFIG INITIALIZED

Configuration storage was reset to factory configuration or erased.

CONV TEMP WARNING

NO2 NO converter temperature outside of warning limits specified by

 

CONV_SET variable.

DATA INITIALIZED

DAS data storage was erased before the last power up occurred.

HVPS WARNING

High voltage power supply output outside of warning limits specified by

 

HVPS_SET variable.

IZS TEMP WARNING 1

IZS temperature outside of warning limits specified by IZS_SET variable.

OZONE FLOW WARNING

Ozone flow outside of warning limits specified by OFLOW_SET variable.

OZONE GEN OFF

Ozone generator is off. This warning message clears itself when the ozone

 

generator is turned on.

PMT TEMP WARNING

PMT temperature outside of warning limits specified by PMT_SET variable.

RCELL PRESS WARN

Reaction cell pressure outside of warning limits specified by

 

RCELL_PRESS_SET variable.

RCELL TEMP WARNING

Reaction cell temperature outside of warning limits specified by RCELL_SET

 

variable.

REAR BOARD NOT DET

Motherboard was not detected during power up.

RELAY BOARD WARN

CPU is unable to communicate with the relay PCA.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SYSTEM RESET

The computer has rebooted.

1 Only Appears if the Internal Span Gas Generator option is installed.

Teledyne Analytical Instruments

91