5.4.3.3.SETUP RNGE MODE AUTO: Configuring the 9110T Analyzer for Auto Range Mode

In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges (LOW and HIGH). The same low and high span settings are applied equally to NO, NO2 and NOX readings.

The unit will switch from LOW range to HIGH range when either the NO, or NOX concentration exceeds 98% of the low range span.

The unit will return from HIGH range back to LOW range once both the NO and NOX concentrations fall below 75% of the low range span.

IMPORTANT

IMPACT ON READINGS OR DATA

 

The LOW & HIGH ranges referred to here are NOT the same as the low &

 

high physical ranges referred to in Section 5.4.1.

Also the RANGE test function displayed on the front panel will be replaced by two separate functions:

RANGE1: The LOW range setting for all analog outputs.

RANGE2: The HIGH range setting for all analog outputs.

The LOW/HIGH range status is also reported through the external, digital status bits (Section 3.3.1.4). To set individual ranges press the following menu sequence.