Page 114
Image 114

Model 9110T NOx Analyzer

Setup Menu

5.9. SETUP DIAG: DIAGNOSTICS FUNCTIONS

A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu. The parameters are dependent on firmware revision (see Appendix A). These tools can be used in a variety of troubleshooting and diagnostic procedures and are referred to in many places of the maintenance and troubleshooting sections of this manual.

The various operating modes available under the DIAG menu are:

Table 5-4: Diagnostic Mode (DIAG) Functions

DIAG SUBMENU

SUBMENU FUNCTION

Front Panel

MANUAL

Mode Indicator

SECTION

 

 

SIGNAL I/O

Allows observation of all digital and analog signals

DIAG I/O

 

in the instrument. Allows certain digital signals such

12.1.3

 

as valves and heaters to be toggled ON and OFF.

 

 

 

 

 

 

ANALOG OUTPUT

When entered, the analyzer performs an analog

 

 

output step test. This can be used to calibrate a

DIAG AOUT

12.7.6.1

 

chart recorder or to test the analog output accuracy.

 

 

 

 

 

 

 

The signal levels of the instruments analog outputs

 

 

ANALOG I/O

may be calibrated (either individually or as a group).

DIAG AIO

 

Various electronic parameters such as signal span,

5.9.2

CONFIGURATION

and offset are available for viewing and

 

 

 

 

 

 

configuration.

 

 

 

 

 

 

TEST CHAN

Selects one of the available test channel signals to

DIAG TCHN

5.9.4

OUTPUT

output over the A4 analog output channel.

 

 

 

 

 

 

 

When activated, the analyzer performs an optic test,

 

 

OPTIC TEST

which turns on an LED located inside the sensor module

DIAG OPTIC

0

near the PMT (Figure 13-20). This diagnostic tests the

 

 

 

 

response of the PMT without having to supply span gas.

 

 

 

 

 

 

ELECTRICAL

When activated, the analyzer performs an electrical test,

 

 

which generates a current intended to simulate the PMT

DIAG ELEC

0

TEST

output to verify the signal handling and conditioning of the

 

 

 

PMT preamp board.

 

 

 

 

 

 

OZONE GEN

Allows the user to manually turn the O3 generator on or

DIAG OZONE

0

off. During initial power up TMR (timer) is displayed while

OVERRIDE 1

the Ozone brick remains off for the first 30 minutes.

 

 

 

 

 

 

 

 

FLOW

This function is used to calibrate the gas flow output

DIAG FCAL

9.7

CALIBRATION 1

signals of sample gas and ozone supply.

 

1 These settings are retained after exiting DIAG mode.

 

 

To access the various DIAG submenus, press the following buttons:

Teledyne Analytical Instruments

114

Page 114