Teledyne 9110T Communication, Configuration Section, Anything other than Multidrop communication

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 49
Image 49

Model 9110TH NOx Analyzer

Getting Started

appropriate cables, each type of communication method, must be configured using the SETUP>COMM menu (see Sections 5.7 and 6).

ETHERNET CONNECTION

For network or Internet communication with the analyzer, connect an Ethernet cable from the analyzer’s rear panel Ethernet interface connector to an Ethernet port. Although the analyzer is shipped with DHCP enabled by default (Section 6.5.2), it should be manually assigned a static IP address.

Configuration: (manual, i.e., static) Section 6.5.1.

USB CONNECTION

 

The USB option can be used for direct communication between the analyzer and a PC; connect a USB

 

cable between the analyzer and computer USB ports. Baud rates must match: check the baud rate on

 

either the computer or the instrument and change the other to match (see Section 6.2.2). This USB

 

connection can only be used when the COM2 port is not in use except for RS-232 Multidrop

 

communication.

 

Configuration: Section 6.6.

Note

If this option is installed, the rear panel COM2 port cannot be used for

 

anything other than Multidrop communication.

RS-232 CONNECTION

For RS-232communications with data terminal equipment (DTE) or with data communication equipment (DCE) connect either a DB9-female-to-DB9-female cable (TAI part number WR000077) or a DB9-female-to-DB25-male cable (Option 60A, Section 1.4), as applicable, from the analyzer’s rear panel RS-232 port to the device. Adjust the DCE-DTE switch (Figure 3-4) to select DTE or DCE as appropriate (Section 6.1).

Configuration: Section 6.3 and Section 6.7.2 (for Hessen protocol).

IMPORTANT

IMPACT ON READINGS OR DATA

 

Cables that appear to be compatible because of matching connectors

 

may incorporate internal wiring that makes the link inoperable. Check

 

cables acquired from sources other than TAI for pin assignments (Figure

 

3-13) before using.

Teledyne Analytical Instruments

49

Page 49
Image 49
Teledyne 9110T Communication, Configuration Section, Anything other than Multidrop communication, Before using