Model 9110T NOx Analyzer

Troubleshooting & Service

Input the starting NO2 concentration value into the 9110T by pressing:

SAMPLE

RANGE=500.0 PPB

NO=XXXX

<TST TST>

CAL

SETUP

Use these buttons to select the appropriate range.

SAMPLE

LOW HIGH

 

 

 

 

 

 

This step only appears if the

 

 

 

 

 

 

 

 

 

 

 

analyzer’s reporting range is

 

 

 

 

 

RANGE TO CAL

 

 

 

set for AUTO range mode.

 

 

 

ENTR EXIT

 

Select LOW and press ENTR.

 

 

 

 

 

 

 

 

Repeat entire procedure for

 

 

 

 

 

 

 

 

 

 

 

 

HIGH range.

 

 

 

 

 

 

 

Repeat entire procedure for each range.

Toggle these buttons to change this value to the concentration of the NO2 gas being used.

SAMPLE

 

RANGE=500.0 PPB

NO=XXXX

 

 

 

<TST

TST>

CAL

 

 

 

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

RANGE=500.0 PPB

NO=XXXX

 

 

<TST TST>

ZERO

 

CONC

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

CONCENTRATON MENU

 

 

 

NOX

NO

CONV

 

 

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Converter Efficiency Menu

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

CONVERTER EFICIENCY MENU

 

 

 

NO2

CAL

SET

 

 

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The expected NO2 span

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

concentration value defaults

M-P CAL

NO2 CE CONC: 500.0 Conc

 

to 400.0 Conc.

 

0

 

4

0

 

 

0.

0

 

0

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

Make sure that you specify

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the actual concentration value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

of the NO2 gas.

 

 

 

 

 

 

 

 

 

 

 

 

 

 