Model 9110TH NOx Analyzer

Calibration Procedures

9.6. CALIBRATION QUALITY ANALYSIS

After completing one of the calibration procedures described above, it is important to evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response curve of the analyzer, separately for NO and NOX. The values for these terms, both individually and relative to each other, indicate the quality of the calibration.

To perform this quality evaluation, you will need to record the values of the following test functions (see Section 4.1.1), all of which are automatically stored in the DAS channel CALDAT for data analysis, documentation and archival.

NO OFFS

NO SLOPE

NOX OFFS

NOX SLOPE

Ensure that these parameters are within the limits listed in Table 9-5 and frequently compare them to those values on the Final Test and Validation Data Sheet (P/N 04490) that came attached to your manual, which should not be significantly different. If they are, refer to the troubleshooting Section 12.

Table 9-5: Calibration Data Quality Evaluation

Function

Minimum Value

Optimum Value

Maximum Value

NOX SLOPE

-0.700

1.000

1.300

NO SLOPE

-0.700

1.000

1.300

NOX OFFS

-20.0 mV

0.0 mV

150.0 mV

NO OFFS

-20.0 mV

0.0 mV

150.0 mV

 

 

 

 

The default DAS configuration records all calibration values in channel CALDAT as well as all calibration check (zero and span) values in its internal memory.

Up to 200 data points are stored for up 4 years of data (on weekly calibration checks) and a lifetime history of monthly calibrations.

Review these data to see if the zero and span responses change over time.

These channels also store the STB figure (standard deviation of NOX concentration) to evaluate if the analyzer response has properly leveled off during the calibration procedure.

Finally, the CALDAT channel also stores the converter efficiency for review and documentation.

Teledyne Analytical Instruments

219

Page 219
Image 219
Teledyne 9110T instruction manual Calibration Quality Analysis, No Offs No Slope NOX Offs NOX Slope