Model 9110T NOx Analyzer

Instrument Maintenance

Figure 11-6: Reaction Cell Assembly

4.Remove two screws (Teledyne P/N SN150) and two washers.

5.The reaction cell will separate into two halves, the stainless steel manifold assembly and the black plastic reaction cell with window gasket, stainless steel reaction cell sleeve, optical filter and O-rings.

6.The reaction cell (both plastic part and stainless steel sleeve) and optical filter should be cleaned with Distilled Water (DI - Water) and a clean tissue, and dried thereafter.

7.Usually it is not necessary to clean the sample and ozone flow orifices since they are protected by sintered filters.

Teledyne Analytical Instruments

255