Model 9110T NOx Analyzer

Troubleshooting & Service

12.3.2.2. Relay PCA Status LEDs

D10 (Green) – NO/NOx Valve

D9 (Green) – AutoZero Valve

D8 (Green) – Optional Sample/Cal Valve

D7 (Green) – Optional Zero/Span Valve

D3 (Yellow) NO2 NO Converter Heater

D2 (Yellow) Reaction Cell Heater

D5 (Yellow) – Optional Internal Span Gas Gen Heater

D11 (Green) – Optional Dual Span Select Valve

D12 (Green) – Optional Pressurized Span Shutoff Valve

D13 (Green) – Optional Pressurized Zero Shutoff Valve

 

 

 

 

D1 (RED)

 

 

 

 

 

Watchdog Indicator

 

 

Figure 12-3: Relay PCA Status LEDS Used for Troubleshooting

 

 

Table 12-5: Relay PCA Status LED Failure Indications

 

 

 

 

 

 

LED

COLOR

FUNCTION

FAULT

INDICATED FAILURE(S)

STATUS

 

 

 

 

 

LED ROW 1

 

 

 

 

 

 

D2

Yellow

Reaction Cell heater

Continuously

Heater broken, thermistor broken

ON or OFF

 

 

 

 

 

D3

Yellow

NO2 converter heater

Continuously

Heater broken, thermocouple broken

ON or OFF

 

 

 

 

 

D7

Green

Zero/Span valve status

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

D8

Green

Sample/Cal valve status

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

D9

Green

Auto-zero valve status

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

D10

Green

NO/NOx valve status

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

LED ROW 2

 

 

 

 

 

 

D5

Yellow

Internal span gas generator

Continuously

Heater broken, thermistor broken

perm tube heater

ON or OFF

 

 

 

 

D11

Green

Dual span select valve

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

D12

Green

Pressurized Span shutoff valve

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

D13

Green

Pressurized Zero shutoff valve

Continuously

Valve broken or stuck, valve driver chip broken

ON or OFF

 

 

 

 

 

D14- 16

Green

Spare

N/A

N/A

 

 

 

 

 

 

 