Model 9110T NOx Analyzer

Setup Menu

EXAMPLE:

NOx Concentration – RANGE1 Set for 0-800 ppb & output A1 set for 0-10 VDC

NO Concentration – RANGE2 Set for 0-200 ppb & output A2 set for 0-5 VDC

NO2 Concentration – RANGE3 Set for 0-400 ppb & output A3 set for 0-5 VDC

Setting analog range limits to different values does not affect the instrument’s calibration.

To select the IND range mode, press the following buttons:

 

SAMPLE

 

RANGE=500.0 PPM

 

NOX= XXXX

 

 

 

 

 

 

 

<TST

TST>

CAL

 

 

 

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Concentration field

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displays all gases.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

 

 

 

 

 

 

 

 

 

CFG

DAS RNGE PASS CLK MORE

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE MODE MENU

 

 

 

 

 

 

 

 

 

 

MODE SET

UNIT

DIL

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

RANGE MODE:SNGL

 

 

 

 

 

 

 

 

 

 

SNGL

IND

AUTO

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

RANGE MODE:IND

 

 

 

 

 

 

 

 

 

 

SNGL

IND

AUTO

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE MODE MENU

 

 

 

 

 

 

 

 

 

 

MODE

SET

UNIT

DIL

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 