Teledyne 9110T Turning AN Analog Output OVER-RANGE Feature ON/OFF, Current Loop Output Check

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 127
Image 127

Model 9110TH NOx Analyzer

Setup Menu

An alternate method for measuring the output of the Current Loop converter is to connect a 250 ohm

±1% resistor across the current loop output in lieu of the current meter (see Figure 3-8 for pin assignments and diagram of the analog output connector). This allows the use of a voltmeter connected across the resistor to measure converter output as VDC or mVDC.

V

+DC Gnd

Figure 5-8: Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels

In this case, follow the procedure above but adjust the output for the following values:

Table 5-8: Current Loop Output Check

% FS

Voltage across

Voltage across

Resistor for 2-20 mA

Resistor for 4-20 mA

 

0

500 mVDC

1000 mVDC

100

5000 mVDC

5000 mVDC

5.9.3.8. TURNING AN ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF

In its default configuration, a ± 5% over-range is available on each of the 9110T’s analog outputs. This over-range can be disabled if your recording device is sensitive to excess voltage or current.

To turn the over-range feature on or off, select the ANALOG I/O CONFIGURATION submenu (see Figure 5-5) then press:

Teledyne Analytical Instruments

127

Page 127
Image 127
Teledyne 9110T instruction manual Turning AN Analog Output OVER-RANGE Feature ON/OFF, Current Loop Output Check