12.2. USING THE ANALOG OUTPUT TEST CHANNEL

The signals available for output over the 9110T’s analog output channel can also be used as diagnostic tools. See Section 5.9.2 for instruction on activating the analog output and selecting a function.

Table 12-3: Test Channel Outputs as Diagnostic Tools

TEST

DESCRIPTION

ZERO

FULL

CAUSES OF EXTREMELY

CHANNEL

SCALE

HIGH / LOW READINGS

 

 

 

 

 

 

Failed PMT

PMT

The output of the PMT

 

 

PMT Temperature too High/Low

0 mV

5000 mV

Bad PMT Preamp PCA

detector converted to a 0

DETECTOR

Failed HVPS

to 5 VDC scale.

 

 

 

 

 

Misadjusted HVPS drive Voltage

 

 

 

 

 

 

 

 

Light Leak in reaction cell

OZONE

The flow rate of O3

0

1000

 

through the analyzer as

Check for Gas Flow problems in the O3 gas lines.

FLOW

measured by the O3 flow

cm3/min

cm3/min

 

sensor

 

 

 

SAMPLE

The calculated flow rate

0

1000

Check for Gas Flow problems in the sample gas lines.

for sample gas through

FLOW

cm3/min

cm3/min

 

the analyzer.

 

 

 

SAMPLE

The pressure of the

 

 

 

sample gas measured

0 In-Hg-A

40 In-Hg-A

Check for Gas Flow problems in the sample gas lines.

PRESSURE

upstream of the Auto

 

 

 

 

Zero Valve

 

 

 

RCELL

The pressure of gas

0 In-Hg-A

40 In-Hg-A

 

inside the reaction cell of

Check for Gas Flow problems in all gas lines.

PRESSURE

the sensor module

 

 

 

 

 

 

 

RCELL TEMP

The temperature of gas

0 °C

70 °C

 

inside the reaction cell of

Same as RCELL TEMP WARNING in Table 12-1.

 

the sensor module

 

 

 

 

The temperature of the

 

 

 

IZS TEMP

permeation tube oven of

0 °C

70 °C

Same as IZS TEMP WARNING in Table 12-1.

the optional internal span

 

gas generator.

 

 

 

CONV TEMP

The temperature NO2

0 mV

5000 mV

Same as CONV TEMP WARNING in Table 12-1.

NO converter

 

 

 

 

PMT TEMP

The temperature inside

0 °C

50 °C

Same as PMT TEMP WARNING in Table 12-1.

PMT

 

 

 

 

BOX TEMP

The temperature inside

0 °C

70 °C

Same as BOX TEMP WARNING in Table 12-1.

the 9110T’s chassis

 

 

 

 

HVPS

Represents the output

0 mV

5000 mV

 

voltage of the PMT's high

Same as HVPSWARNING in Table 12-1.

VOLTAGE

voltage power supply

 

 

 

 

 

 

 

 

 

 

 

 