Teledyne 9110T Cylinders, Available, Calibration Span GAS, NIST-SRM Type Nominal Concentration

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 57
Image 57

Model 9110TH NOx Analyzer

Getting Started

Pure nitrogen (N2) could be used as a zero gas for applications where NOX is measured in nitrogen.

If your analyzer is equipped with an external zero air scrubber option, it is capable of creating zero air from ambient air.

For analyzers without the external zero air scrubber, a zero air generator such as the Teledyne Model 701 can be used. Please visit the company website for more information.

CALIBRATION (SPAN) GAS

Calibration gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired reporting range. To measure NOX with the 9110T NOX analyzer, it is recommended that you use a span gas with an NO concentration equal to 80% of the measurement range for your application

EXAMPLE:

If the application is to measure NOX in ambient air between 0 ppm and 500 ppb, an appropriate span gas would be 400 ppb.

If the application is to measure NOX in ambient air between 0 ppm and 1000 ppb, an appropriate span gas would be 800 ppb.

Even though NO gas in nitrogen could be used as a span gas, the matrix of the balance gas is different and may cause interference problems or yield incorrect calibrations.

The same applies to gases that contain high concentrations of other compounds (for example, CO2 or H2O).

The span gas should match all concentrations of all gases of the measured medium as closely as possible.

Cylinders of

the following types of calibrated NOx and NO gas traceable to NIST standards specifications

(also referred to as EPA protocol calibration gases or Standard Reference Materials) are commercially

available.

 

 

 

 

 

 

 

 

Table 3-8:

NIST-SRM's Available for Traceability of NOx Calibration Gases

 

 

 

 

 

 

 

 

 

NIST-SRM4

 

TYPE

NOMINAL

 

 

 

 

 

 

CONCENTRATION

 

 

 

2627a

 

Nitric Oxide (NO) in N2

5 ppm

 

 

 

2628a

 

Nitric Oxide (NO) in N2

10 ppm

 

 

 

2629a

 

Nitric Oxide (NO) in N2

20 ppm

 

 

 

1683b

 

Nitric Oxide (NO) in N2

50 ppm

 

 

 

1684b

 

Nitric Oxide (NO) in N2

100 ppm

 

 

 

1685b

 

Nitric Oxide (NO) in N2

250 ppm

 

 

 

1686b

 

Nitric Oxide (NO) in N2

5000 ppm

 

 

 

1687b

 

Nitric Oxide (NO) in N2

1000 ppm

 

 

 

2630

 

Nitric Oxide (NO) in N2

1500 ppm

 

 

 

2631a

 

Nitric Oxide (NO) in N2

3000 ppm

 

 

 

2635

 

Nitric Oxide (NO) in N2

800 ppm

 

 

 

2636a

 

Nitric Oxide (NO) in N2

2000 ppm

 

 

 

2631a

 

Oxides of Nitrogen (NOx) in N2

2500 ppm

 

 

 

1684b

 

Oxides of Nitrogen (NOx) in N2

100 ppm

 

Note

 

The NO2 permeation tube included with the 9110T’s optional Internal

Teledyne Analytical Instruments

57

Page 57
Image 57
Teledyne 9110T instruction manual Cylinders, Available, Calibration Span GAS, NIST-SRM Type Nominal Concentration