Model 9110T NOx Analyzer

Specifications, Approvals, & Compliance

 

 

 

 

Parameter

Specification

 

 

External Pump Pack: 15 lbs (7 kg)

 

Operating Temperature Range

5 - 40 °C (with EPA equivalency)

 

Humidity Range

0-95% RH non-condensing

 

 

Installation Category (Over voltage Category) II Pollution Degree 2

 

Environmental Conditions

1As defined by the US EPA.

2Defined as twice the zero noise level by the US EPA.

2.2. EPA EQUIVALENCY DESIGNATION

TAI’s 9110T nitrogen oxides analyzer is designated as a reference method (Number RFNA-1104-099) for NO2 measurement, as defined in 40 CFR Part 53, when operated under the following conditions:

Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).

Ambient temperature range of 5 to 40°C.

With 1-micron PTFE filter element installed in the internal filter assembly.

Equipped with ozone supply air filter

Gas flow supplied by External vacuum pump capable of 10 in-Hg-A at 2 standard liters per minute (slpm) or better.

Software Settings, see Table 2-2:

Table 2-2: Software Settings for EPA Equivalence

Parameter

Setting

Dynamic Zero

OFF or ON

 

 

Dynamic Span

OFF

 

 

CAL-on-NO2

OFF

 

 

Dilution Factor

1.0 or OFF

Temp/Pres compensation

ON

AutoCal

ON or OFF

Independent range

ON or OFF

Auto range

ON or OFF

Under the designation, the Analyzer may be operated with or without the following options:

Rack mount with or without slides.

Rack mount for external pump.

4-20mA isolated analog outputs.

Zero/Span Valves option.

Internal Zero/Span (IZS) option with:

NO2 permeation tube - 0.4 ppm at 0.7 liter per minute; certified/uncertified.

NO2 permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.

Teledyne Analytical Instruments

28

Page 28
Image 28
Teledyne 9110T EPA Equivalency Designation, Parameter Specification, Software Settings for EPA Equivalence Parameter