Teledyne 9110T instruction manual OUT-GASSING no NOX, NOx Loss

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 299
Image 299

 

Model 9110T NOx Analyzer

 

Troubleshooting & Service

 

12.7.11.1. Simplified GPT Data Sheet

 

 

 

Line # TEST

 

RESULT

 

1

LEAK-CHECK (WHEN HOT)

 

YES / NO

2

NOx RESPONSE (MOLY BYPASSED)

__________

 

3

NOx RESPONSE (MOLY IN-LINE)

 

__________

 

4

OUT-GASSING (NO – NOX)

 

__________ (>-5 ppb, <5 ppb)

5

(NOx ORIG)

(NOx mode, O3 off)

__________ ppb

6

(NOx REM)

(NOx mode, O3 on)

__________ ppb

7

NOx LOSS

 

__________ (A)

 

 

 

 

(<4% of NOx ORIG:

 

 

 

 

for 450PP 4% is 18 ppb)

8

(NO ORIG)

(NO mode, O3 off)

__________ ppb

9

(NO REM)

(NO mode, O3 on)

__________ ppb

10

NO2

__________ (B) (>300ppb)

11Efficiency LOSS [ ( A / B ) x 100 ] = [ ( ____A / ____B ) x 100 ] = ____C%

12Total Conv Eff [ 100% – C ] = 100% - ____C = _____D% ( > 96%)

Teledyne Analytical Instruments