Model 9110TH NOx Analyzer

Getting Started

Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer and;

The pneumatic setup matches that described in Section 3.3.2.

VERIFYING THE REPORTING RANGE SETTINGS

While it is possible to perform the following procedure with any range setting we recommend that you perform this initial checkout using following reporting range settings:

Unit of Measure: PPB

Reporting Range: 500 ppb

Mode Setting: SNGL

While these are the default setting for the 9110T analyzer, it is recommended that you verify them before proceeding with the calibration procedure, by pressing the following menu button sequence:

Teledyne Analytical Instruments

79

Page 79
Image 79
Teledyne 9110T instruction manual Verifying the Reporting Range Settings