Teledyne 9110T Predictive Diagnostics, Maintenance Procedures, Predictive Uses for Test Functions

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 241
Image 241

Model 9110T NOx Analyzer

Instrument Maintenance

11.2. PREDICTIVE DIAGNOSTICS

Predictive diagnostic functions including failure warnings and alarms built into the analyzer’s firmware allow the user to determine when repairs are necessary.

The Test Functions can also be used to predict failures by looking at how their values change over time. Initially it may be useful to compare the state of these Test Functions to the values recorded on the printed record of the Final Test and Validation Data Sheet, P/N 04490.

The following table can be used as a basis for taking action as these values change with time. The internal data acquisition system (DAS) is a convenient way to record and track these changes. Use APICOM (Section 8.1.1) to download and review this data from a remote location.

Table 11-2: Predictive Uses for Test Functions

FUNCTION

EXPECTED

ACTUAL

INTERPRETATION & ACTION

RCEL

Constant to within

Fluctuating

Developing leak in pneumatic system. Check for leaks.

 

 

 

Pump performance is degrading. Rebuild pump when pressure

(pressure)

± 0.5 in-Hg-A

Slowly increasing

 

 

 

is above 10 in-Hg-A.

 

Constant within

Fluctuating

Developing leak in pneumatic system. Check for leaks.

SAMP

Slowly increasing

Flow path is clogging up. Replace orifice filters.

(pressure)

atmospheric

 

 

changes

Slowly decreasing

Developing leak in pneumatic system to vacuum (developing

 

 

valve failure). Check for leaks.

 

 

 

OZONE FL

Constant to within

Slowly decreasing

Flow path is clogging up. Replace orifice filters.

± 15

 

 

 

 

 

 

Developing AZERO valve failure. Replace valve.

 

Constant within

 

 

AZERO

Significantly increasing

PMT cooler failure. Check cooler, circuit, and power supplies.

±20 of check-out

 

Developing light leak.

 

value

 

 

 

 

O3 air filter cartridge is exhausted. Change chemical.

NO2

Constant for

Slowly decreasing

Converter efficiency may be degrading. Replace converter

constant

signal for same

(Concentration)

components.

concentrations

concentration

 

 

 

 

 

 

NO2

 

 

Change in instrument response. Low level (hardware) calibrate

Constant

Decreasing over time

the sensor.

with IZS Option

 

 

response from day

 

Degradation of IZS permeation tube. Change permeation tube.

installed

 

to day

 

 

Heavily fluctuating from

Ambient changes in moisture are affecting the performance.

(Concentration)

 

 

 

day to day

Add a dryer to the zero air inlet.

NO

Constant for

 

Drift of instrument response; clean RCEL window. Check for

constant

Decreasing over time

(Concentration)

flow leaks or irregularities.

concentration

 

 

 

 

 

 

 

 

Note

It is recommended that the above functions be checked weekly.

11.3. MAINTENANCE PROCEDURES

The following procedures are to be performed periodically as part of the standard maintenance of the 9110T.

Teledyne Analytical Instruments

241

Page 241
Image 241
Teledyne 9110T instruction manual Predictive Diagnostics, Maintenance Procedures, Predictive Uses for Test Functions, Samp