Model 9110TH NOx Analyzer

Getting Started

3.4.3. FUNCTIONAL CHECKS

After the analyzer’s components have warmed up for at least 60 minutes, verify that the software properly supports any hardware options that are installed.

For information on navigating through the analyzer’s software menus, see the menu trees described in Appendix A.1.

Check to ensure that the analyzer is functioning within allowable operating parameters.

Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as their expected values.

These functions are also useful tools for diagnosing problems with your analyzer.

The enclosed Final Test and Validation Data sheet (P/N 04409) lists these values before the instrument left the factory.

To view the current values of these parameters press the following button sequence on the analyzer’s front panel. Remember until the unit has completed its warm up these parameters may not have stabilized.

3.4.4. INITIAL CALIBRATION

To perform the following calibration you must have sources for zero air and calibration (span) gas available for input into the inlet/outlet fittings on the back of the analyzer (see Section 3.3.2.1).

Teledyne Analytical Instruments

77

Page 77
Image 77
Teledyne 9110T instruction manual Functional Checks, Initial Calibration