Model 9110T NOx Analyzer

Calibration Procedures

9.5.1. SETUP ACAL: PROGRAMMING AND AUTO CAL SEQUENCE

To program the example sequence shown in Table 9-4, press:

SAMPLE

RANGE = 500.0 PPB

NOX=XXX.X

< TST TST > CAL CALZ CZLS

SETUP

 

 

 

SETUP X.X

CFG ACAL DAS RNGE PASS CLK MORE EXIT

SETUP X.X SEQ 1) DISABLED

 

 

NEXT MODE

 

 

EXIT

 

 

 

 

SETUP X.X SEQ 2) DISABLED

 

PREV NEXT MODE

 

 

EXIT

 

 

 

 

 

 

SETUP X.X

MODE: DISABLED

 

 

NEXT

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

SETUP X.X

MODE: ZERO

 

 

PREV

NEXT

 

 

 

ENTR EXIT

 

 

 

 

 

 

SETUP X.X

MODE: ZERO–SPAN

 

PREV

NEXT

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

SETUP X.X

SEQ 2) ZERO–SPAN, 1:00:00

PREV NEXT MODE

SET

 

EXIT

 

 

 

 

SETUP X.X TIMER ENABLE: ON

 

 

SET> EDIT

 

 

EXIT

 

 

SETUP X.X STARTING DATE: 01–JAN–07

<SET SET>

EDIT

 

 

EXIT

 

 

 

 

SETUP X.X

STARTING DATE: 01–JAN–02

0

4

SEP

0

8

ENTR EXIT

 

 

 

 

 

 

Toggle buttons to set

Day, Month & Year:

Format : DD-MON-YY

CONTINUE NEXT PAGE

With STARTING TIME

Teledyne Analytical Instruments

216

Page 216
Image 216
Teledyne 9110T Setup  Acal Programming and Auto CAL Sequence, To program the example sequence shown in -4, press