Teledyne 9110T Setup for Calibration with the Internal Span GAS Generator, CAL on NO2 Feature

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 203
Image 203

Model 9110TH NOx Analyzer

Calibration Procedures

9.3.2.SETUP FOR CALIBRATION WITH THE INTERNAL SPAN GAS GENERATOR

Connect the sources of zero air and span gas as shown in Figure 9-4.

Figure 9-4: Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas Generator

9.3.3. CAL ON NO2 FEATURE

When using the IZS option to calibrate the 9110T, the analyzer’s CAL_ON_NO2 feature must be turned on. This feature enables a continuous zero gas flow across the IZS permeation tube and through the NO2 converter. It also programs the analyzer to use the NO output from the NO2 converter to calibrate the span value of both NO and NOX.

Note

This feature should only be enabled when a span calibration or calibration

 

check is performed.

While CAL_ON_NO2 is enabled, the NO2 concentration will always be reported as zero. This is because the gas is continuously routed through the NO2 converter and the analyzer’s firmware simulates calibration with NO gas.

Table 9-1: IZS Option Valve States with CAL_ON_NO2 Turned ON

Valve

Condition

Valve Port Connections

Sample/Cal

Open to zero/span valve

1 2

 

 

 

Zero/Span

Open to SPAN GAS inlet

1 2

NO/NOx Valve

Open to NO2 converter

1 2

 

 

 

Auto Zero Valve

Cycles normally

N/A

Since the instrument sees the same concentration of NO during both NO and NOX cycles, it reports an NO2 concentration of zero.

TO turn the CAL_ON_NO2 feature ON/OFF, press:

Teledyne Analytical Instruments

203

Page 203
Image 203
Teledyne 9110T Setup for Calibration with the Internal Span GAS Generator, CAL on NO2 Feature, Check is performed