Teledyne 9110T Gas Conditioner Options, Ammonia Removal Sample Conditioner OPT 86A

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 72
Image 72

Model 9110T NOx Analyzer

 

 

Getting Started

 

 

 

 

 

 

 

 

Zero/Span

Open to ZERO AIR inlet

3 2

 

 

ZERO CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

 

 

Zero/Span

Open to ZERO AIR inlet

3 2

 

 

 

 

 

SPAN CAL

Sample/Cal

Open to ZERO/SPAN Valve

1 2

 

 

Zero/Span

Open to SPAN inlet

1 2

 

 

 

 

3.3.2.6. Gas Conditioner Options

AMMONIA REMOVAL SAMPLE CONDITIONER (OPT 86A)

The 9110T includes a Nafion® permeation gas exchange tube to remove H2O a from the ozone generator supply gas stream to a dew point of about -20° C (~600 ppm H2O) and effectively remove concentrations of ammonia (NH3) up to about 1 ppm.

An additional Sample Conditioner can be added to the 9110T’s sample gas stream.

Sample Gas

 

NH3 Dryer/Scrubber

 

NO/NOX

 

VALVE

 

SAMPLE

VACUUM

PRESSURE

PRESSURE

SENSOR

SENSOR

 

O3 FLOW

 

SENSOR

AUTOZERO

 

VALVE

 

EXHAUST MANIFOLD

 

PMT

 

PERMAPURE DRYER

 

Figure 3-28: Pneumatics for Sample Conditioner OPT 86A

ZERO AIR SCRUBBER (OPT 86C), FOR Z/S VALVES