Teledyne 9110T instruction manual Zero Calibration, Span Calibration

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 230
Image 230

Model 9110T NOx Analyzer

EPA Protocol Calibration

IMPORTANT

IMPACT ON READINGS OR DATA

The analyzer should be calibrated on the same range used for monitoring.

Ensure that the GPT calibration system can supply the range of concentrations at a sufficient flow over the whole range of concentrations that will be encountered during calibration.

10.6.2. ZERO CALIBRATION

The zero point calibration procedure is identical to that described in Sections 9.1, 9.2, 9.4 or 9.5.

10.6.3.SPAN CALIBRATION

1.Adjust the NO concentration to approximately 80% of the URL of the NO channel.

The expected NO and NOX span concentrations can be determined by measuring the cylinder and diluent flows and computing the resulting concentrations.

If there is any NO2 impurity in the NO standard gas it should be taken into account when the NOX concentration is entered during the NO/NOX channel calibration.

This is done by ADDING the impurity concentration to the NO concentration to get the NOX concentration for calibration. Calculate the exact NO and NOX concentrations as follows:

Equation 10-9

[NO ]OUT = FNO x [NO ]STD

FT

2.Enter the respective concentrations using the procedure in Section 4.2. The expected span concentrations need not be re-entered each time a calibration is performed unless they are changed.

3.Enter the expected NOX and NO span gas concentrations:

4.Sample the generated concentration until the NO and the NOX responses have stabilized.

5.Span the instrument by the following the same method as 9.1, 9.3 or 9.5.

The analog voltage output should measure 80% of the voltage range selected. (e.g.

4.00VDC if 0-5V output is selected.)

The readings on the front panel display should be equal to the expected NO and NOX concentrations.

Teledyne Analytical Instruments

230

Page 230
Image 230
Teledyne 9110T instruction manual Zero Calibration, Span Calibration