Teledyne 9110T instruction manual Default Ethernet setting is Dhcp

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 143
Image 143

Model 9110TH NOx Analyzer

Communications Setup and Operation

 

Table 6-3. LAN/Ethernet Default Configuration Properties

 

 

 

 

 

 

PROPERTY

DEFAULT STATE

DESCRIPTION

 

 

DHCP

 

This displays whether the DHCP is turned ON or OFF. Press

 

 

ON

EDIT and toggle ON for automatic configuration after first

 

 

 

 

consulting network administrator.

 

 

INSTRUMENT

 

This string of four packets of 1 to 3 numbers each (e.g.

 

 

IP ADDRESS

 

192.168.76.55.) is the address of the analyzer itself.

 

 

GATEWAY IP

0.0.0.0

Can only be edited when DHCP is set to OFF.

 

 

A string of numbers very similar to the Instrument IP address

 

 

ADDRESS

 

(e.g. 192.168.76.1.) that is the address of the computer used by

 

 

 

 

your LAN to access the Internet.

 

 

 

 

Can only be edited when DHCP is set to OFF.

 

 

 

 

Also a string of four packets of 1 to 3 numbers each (e.g.

 

 

 

0.0.0.0

255.255.252.0) that identifies the LAN to which the device is

 

 

SUBNET MASK

connected.

 

 

 

All addressable devices and computers on a LAN must have the

 

 

 

 

 

 

 

 

same subnet mask. Any transmissions sent to devices with

 

 

 

 

different subnets are assumed to be outside of the LAN and are

 

 

 

 

routed through the gateway computer onto the Internet.

 

 

TCP PORT1

3000

This number defines the terminal control port by which the

 

 

instrument is addressed by terminal emulation software, such as

 

 

 

 

 

 

 

Internet or Teledyne API’s APICOM.

 

 

 

[initially blank]

The name by which your analyzer will appear when addressed

 

 

HOST NAME

from other computers on the LAN or via the Internet. To assign

 

 

 

 

 

 

 

or change, see Section 6.5.2.1.

 

 

 

 

 

 

 

1 Do not change the setting for this property unless instructed to by TAI’s Customer Service personnel.

 

6.5.2.CONFIGURING ETHERNET COMMUNICATION USING DYNAMIC HOST CONFIGURATION PROTOCOL (DHCP)

The default Ethernet setting is DHCP.

Teledyne Analytical Instruments

143

Page 143
Image 143
Teledyne 9110T instruction manual Default Ethernet setting is Dhcp