Model 9110T NOx Analyzer

 

Setup Menu

 

 

 

 

Table 5-3: Variable Names (VARS)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ALLOWED

VARS

 

 

NO.

VARIABLE

 

DESCRIPTION

DEFAULT

 

 

 

 

 

 

VALUES

VALUES

 

 

 

 

 

 

 

 

 

 

 

 

Changes the Internal Data Acquisition System (DAS)

May be set for

 

 

 

 

 

 

HOLDOFF timer:

 

 

 

 

 

 

intervals

 

 

 

0

 

 

No data is stored in the DAS channels during situations

 

 

 

DAS_HOLD_OFF

 

between

15 min.

 

 

 

when the software considers the data to be questionable

 

 

 

 

 

0.5 – 20 min

 

 

 

 

 

 

such as during warm up of just after the instrument returns

 

 

 

 

 

 

 

 

 

 

 

 

 

from one of its calibration mode to SAMPLE Mode.

 

 

 

 

1

 

 

Selects the gas measurement mode in which the

NO;

NOx–NO

 

 

MEASURE_MODE

 

instrument is to operate. NOx only, NO only or NOx and NO

NOx;

 

 

 

 

 

simultaneously.

NOx–NO

 

 

 

2

STABIL_GAS

 

Selects which gas measurement is displayed when the

NO; NOx;

NOx

 

 

 

STABIL test function is selected

NO2;

 

 

3

TPC_ENABLE

 

Enables or disables the Temperature and Pressure

ON/OFF

ON

 

 

 

Compensation (TPC) feature (Section 13.9.2).

 

 

4

DYN_ZERO 1

 

Dynamic zero automatically adjusts offset and slope of the

 

OFF

 

 

 

NO and NOX response when performing a zero point

ON/OFF

 

 

 

 

 

calibration during an AutoCal (see Section 9.5).

 

 

 

 

5

DYN_SPAN 1

 

Dynamic span automatically adjusts the offsets and slopes

 

OFF

 

 

 

of the NO and NOx response when performing a sp point

ON/OFF

 

 

 

 

 

calibration during an AutoCal (see Section 9.5).

 

 

 

 

6

IZS_SET

 

Sets the internal span gas generator’s permeation tube

 

51 C

 

 

 

 

oven temperature. Changing this temperature will impact

30°C - 70°C

°

 

 

 

 

 

the NO2 permeation rate (Section 3.3.2.5).

 

 

 

 

7

CONC_PRECISION

Allows the user to set the number of significant digits to the

AUTO, 1, 2,

AUTO

 

 

right of the decimal point display of concentration and

 

 

3, 4

 

 

 

 

 

stability values.

 

 

 

 

8

 

 

Adjusts the speed of the analyzer’s clock. Choose the +

-60 to +60

0 sec

 

 

CLOCK_ADJ

 

sign if the clock is too slow, choose the - sign if the clock is

 

 

 

s/day

 

 

 

 

 

too fast.

 

 

 

 

 

 

 

Allows turning ON and OFF the ability to span the analyzer

 

 

 

 

9

CAL_ON_NO2

 

with NO2, in which case the instrument acts as if NO and

ON or OFF

OFF

 

 

 

NOX are spanned, even though it is supplied with NO2.

 

 

 

 

 

The NO2 concentration is then zero by default.

 

 

 

 

 

 

 

 

 

 

 

1 Use of the DYN_ZERO and DYN_SPAN features are not allowed for applications requiring EPA equivalency.

 

 

 

IMPORTANT

 

IMPACT ON READINGS OR DATA

 

 

 

 

 

There are more VARS available when using the password, 929, for

 

 

 

configuration. Use caution when pressing any buttons while in this

 

 

 

setup. Any changes made may alter the performance of the instrument

 

 

 

or cause the instrument to not function properly. Note that if there is an

 

 

 

accidental change to a setup parameter, press EXIT to discard the

 

 

 

changes.

 

 

 

 

Note

There is a 2-sec latency period between when a VARS value is changed

 

 

and the new value is stored into the analyzer’s memory. DO NOT turn

 

 

the analyzer off during this period or the new setting will be lost.

Teledyne Analytical Instruments

112