Model 9110TH NOx Analyzer

Setup Menu

5.9.3.6. Manual Calibration of the Analog Outputs Configured for Voltage Ranges

For highest accuracy, the voltages of the analog outputs can be manually calibrated.

Note

The menu for manually adjusting the analog output signal level will only

 

appear if the AUTO-CAL feature is turned off for the channel being

 

adjusted. (See Section 5.9.3.3).

Calibration is performed with a voltmeter connected across the output terminals and by changing the actual output signal level using the front panel buttons in 100, 10 or 1 count increments. See Figure 3-8 for pin assignments and diagram of the analog output connector.

V

+DC Gnd

Figure 5-6: Setup for Checking / Calibrating DCV Analog Output Signal Levels

Table 5-7: Voltage Tolerances for the TEST CHANNEL Calibration

FULL

ZERO

 

SPAN

MINIMUM

SPAN VOLTAGE

ADJUSTMENT

SCALE

TOLERANCE

TOLERANCE

 

(1 count)

 

 

 

 

0.1 VDC

±0.0005V

90 mV

±0.001V

0.02 mV

1 VDC

±0.001V

900 mV

±0.001V

0.24 mV

 

 

 

 

 

5 VDC

±0.002V

4500 mV

±0.003V

1.22 mV

 

 

 

 

 

10 VDC

±0.004V

4500 mV

±0.006V

2.44 mV

 

 

 

 

 

Teledyne Analytical Instruments

123

Page 123
Image 123
Teledyne 9110T instruction manual Adjusted. See Section