Model 9110T NOx Analyzer

Setup Menu

5.9.4.TEST CHAN OUTPUT (SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4)

The test functions available to be reported are listed in Table 5-9:

Table 5-9: Test Channels Functions available on the 9110T’s Analog Output

TEST CHANNEL

DESCRIPTION

ZERO

FULL SCALE

 

 

 

 

NONE

TEST CHANNEL IS TURNED OFF

 

PMT DETECTOR

The output of the PMT detector

0 mV

5000 mV 1

 

converted to a 0 to 5 VDC scale.

 

 

OZONE FLOW

The flow rate of O3 through the analyzer

0 cm3/min

1000 cm3/min

 

as measured by the O3 flow sensor.

 

 

SAMPLE FLOW

The calculated flow rate for sample gas

0 cm3/min

1000 cm3/min

 

through the analyzer.

 

 

 

The pressure of the sample gas

 

 

SAMPLE PRESSURE

measured upstream of the Auto Zero

0 Hg-In-A

40 "Hg-In-A

 

Valve.

 

 

RCELL PRESSURE

The pressure of gas inside the reaction

0 Hg-In-A

40 Hg-In-A

cell of the sensor module.

 

 

 

RCELL TEMP

The temperature of gas inside the

0 °C

70 °C

reaction cell of the sensor module.

 

 

 

MANIFOLD TEMP

Not used in the Model 9110T.

 

 

 

The temperature of the permeation tube

 

 

IZS TEMP

oven of the optional internal span gas

0 °C

70 °C

 

generator.

 

 

CONV TEMP

The temperature NO2 NO converter.

0 °C

500 °C

PMT TEMP

The temperature inside PMT.

0 °C

50 °C

BOX TEMP

The temperature inside the 9110T’s

0 °C

70 °C

chassis.

 

 

 

HVPS VOLTAGE

Represents the output voltage of the

0 mV

5000 mV 1

 

PMT's high voltage power supply.

 

 

Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds TEST to the list of test functions viewable via the front panel display.