Model 9110T NOx Analyzer

Setup Menu

5.9.4.TEST CHAN OUTPUT (SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4)

The test functions available to be reported are listed in Table 5-9:

Table 5-9: Test Channels Functions available on the 9110T’s Analog Output

TEST CHANNEL

DESCRIPTION

ZERO

FULL SCALE

 

 

 

 

NONE

TEST CHANNEL IS TURNED OFF

 

PMT DETECTOR

The output of the PMT detector

0 mV

5000 mV 1

 

converted to a 0 to 5 VDC scale.

 

 

OZONE FLOW

The flow rate of O3 through the analyzer

0 cm3/min

1000 cm3/min

 

as measured by the O3 flow sensor.

 

 

SAMPLE FLOW

The calculated flow rate for sample gas

0 cm3/min

1000 cm3/min

 

through the analyzer.

 

 

 

The pressure of the sample gas

 

 

SAMPLE PRESSURE

measured upstream of the Auto Zero

0 Hg-In-A

40 "Hg-In-A

 

Valve.

 

 

RCELL PRESSURE

The pressure of gas inside the reaction

0 Hg-In-A

40 Hg-In-A

cell of the sensor module.

 

 

 

RCELL TEMP

The temperature of gas inside the

0 °C

70 °C

reaction cell of the sensor module.

 

 

 

MANIFOLD TEMP

Not used in the Model 9110T.

 

 

 

The temperature of the permeation tube

 

 

IZS TEMP

oven of the optional internal span gas

0 °C

70 °C

 

generator.

 

 

CONV TEMP

The temperature NO2 NO converter.

0 °C

500 °C

PMT TEMP

The temperature inside PMT.

0 °C

50 °C

BOX TEMP

The temperature inside the 9110T’s

0 °C

70 °C

chassis.

 

 

 

HVPS VOLTAGE

Represents the output voltage of the

0 mV

5000 mV 1

 

PMT's high voltage power supply.

 

 

Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds TEST to the list of test functions viewable via the front panel display.

Teledyne Analytical Instruments

132

Page 132
Image 132
Teledyne 9110T instruction manual Test functions available to be reported are listed in Table