7.2.2. DAS CONFIGURATION VIA TERMINAL EMULATION PROGRAMS

Although TAI recommends the use of APICOM, the DAS can also be accessed and configured through a terminal emulation program such as HyperTerminal (see Figure 7-4 for example). It is best to start by downloading the default DAS configuration, getting familiar with its command structure and syntax conventions, and then altering a copy of the original file offline before uploading the new configuration.

Figure 7-4: DAS Configuration Through a Terminal Emulation Program

See Section 8.2.1 for configuration commands and their strict syntax. Commands can be pasted in from of an existing text file, which was first edited offline and then uploaded through a specific transfer procedure.

IMPORTANT

IMPACT ON READINGS OR DATA

 

Whereas the editing, adding and deleting of DAS channels and

 

parameters of one channel through the front-panel control buttons can

 

be done without affecting the other channels, uploading a DAS

 

configuration script to the analyzer through its communication ports will

 

erase all data, parameters and channels by replacing them with the new

 

DAS configuration. Backup of data and the original DAS configuration is

 

advised before attempting any DAS changes.