Model 9110T NOx Analyzer

Setup Menu

From the

DIAG

 

 

 

PREV

NEXT

EXIT

 

<SET

CAL

EXIT

 

Continue pressing

until you reach the

 

 

output to be configured

 

<SET

 

EXIT

 

Firmware automatically performs a zero point calibration

 

 

of the Motherboard’s analog Inputs

 

Firmware automatically performs a span point calibration

 

 

of the Motherboard’s analog Inputs

 

Perform

 

 

 

Troubleshooting or call

 

 

 

Teledyne API’s

 

<SET

EXIT

Customer Service

 

 

 

5.9.3.11. External Analog Inputs (XIN1…XIN8) Option Configuration

To configure the analyzer’s optional external analog inputs, define for each channel:

gain (number of units represented by 1 volt)

offset (volts)

engineering units to be represented in volts (each press of the touchscreen button scrolls the list of alphanumeric characters from A-Z and 0-9)

whether to display the channel in the Test functions