Model 9110TH NOx Analyzer

Specifications, Approvals, & Compliance

2. SPECIFICATIONS, APPROVALS, & COMPLIANCE

This section presents specifications for the 9110T, Agency approvals, EPA designation, and CE mark and safety compliance..

2.1. SPECIFICATIONS

Table 2-1 presents the instrument’s parameters and the specifications that each meets.

Table 2-1: 9110T Basic Unit Specifications

Parameter

Min/Max Range (Physical Analog Output)

Measurement Units

Zero Noise1

Span Noise1

Lower Detectable Limit 2

Zero Drift

Span Drift

Lag Time1

Rise/Fall Time1

Linearity

Precision

Sample Flow Rate

AC Power

Power, Ext Pump

Analog Output Ranges

Analog Output Resolution

Recorder Offset

Standard I/O

Optional I/O

Dimensions H x W x D

Weight

Specification

Min: 0-50 ppb Full Scale

Max: 0-20,000 ppb Full Scale (selectable, independent NO, NO2, NOx ranges and auto ranges supported)

ppb, ppm, µg/m3, mg/m3 (selectable)

<0.2 ppb (RMS)

<0.5% of reading (RMS) above 50 ppb or 0.2 ppb, whichever is greater 0.4 ppb

<0.5 ppb (at constant temperature and voltage) /24 hours

<0.5% of Full Scale (at constant temperature and voltage) /24 hours 20 seconds

<60 seconds to 95%

1% of full scale / 24 hours

0.5% of reading above 50 ppb

500 cm3/min ± 10%

100-120 V, 60 Hz (3.0 A); 220-240 V, 50 Hz (2.5 A)

100 V, 50/60 Hz (3.25 A); 115 V, 60 Hz (3.0 A); 220-240 V, 50/60 Hz (2.5 A)

10V, 5V, 1V, 0.1V (selectable)

All Ranges with 5% Under/Over Range

1 part in 4096 of selected full-scale voltage

±10%

1 Ethernet: 10/100Base-T

2 RS-232 (300 – 115,200 baud)

2 USB device ports

8 opto-isolated digital status outputs (7 defined, 1 spare) 6 opto-isolated digital control inputs (4 defined, 2 spar)

4 analog outputs

1 USB com port

1 RS485

8 analog inputs (0-10V, 12-bit)

4 digital alarm outputs Multidrop RS232

3 4-20mA current outputs

7" x 17" x 23.5" (178mm x 432 mm x 597 mm) Analyzer: 40 lbs (18 kg)

Teledyne Analytical Instruments

27

Page 27
Image 27
Teledyne instruction manual 9110T Basic Unit Specifications Parameter