Teledyne 9110T instruction manual Other Performance Problems, Excessive Noise, Slow Response

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 281
Image 281

Model 9110T NOx Analyzer

Troubleshooting & Service

The converter efficiency parameter has been set to a value not equal to 1.000 even though the conversion efficiency is 1.0.

The actual conversion efficiency needs to match the parameter set in the CAL menu.

See Section 9.1.4 for more information on this feature.

An instrument calibration with the IZS option (and expected concentrations set to the same amount) will always yield identical slopes for NO and NOX, as the instrument measures only NOX and assumes NO to be the same (with NO2 being zero).

12.6. OTHER PERFORMANCE PROBLEMS

Dynamic problems (i.e. problems that only manifest themselves when the analyzer is monitoring sample gas) can be the most difficult and time consuming to isolate and resolve. The following section provides an itemized list of the most common dynamic problems with recommended troubleshooting checks and corrective actions.

12.6.1. EXCESSIVE NOISE

Excessive noise levels under normal operation usually indicate leaks in the sample supply or the analyzer itself. Ensure that the sample or span gas supply is leak-free and carry out a detailed leak check as described earlier in this chapter.

Another possibility of excessive signal noise may be the preamplifier board, the high voltage power supply and/or the PMT detector itself.

Contact the factory on troubleshooting these components.

12.6.2. SLOW RESPONSE

If the analyzer starts responding too slow to any changes in sample, zero or span gas, check for the following:

Dirty or plugged sample filter or sample lines.

Sample inlet line is too long.

Leaking NO/NOX valve. Carry out a leak check.

Dirty or plugged critical flow orifices. Check flows, pressures and, if necessary, change orifices (Section 11.3.10).

Wrong materials in contact with sample - use glass, stainless steel or Teflon materials only. Porous materials, in particular, will cause memory effects and slow changes in response.

Dirty reaction cell. Clean the reaction cell.

Insufficient time allowed for purging of lines upstream of the analyzer. Wait until stability is low.

Insufficient time allowed for NO or NO2 calibration gas source to become stable. Wait until stability is low.

NO2 converter temperature is too low. Check for proper temperature.

Teledyne Analytical Instruments

281

Page 281
Image 281
Teledyne 9110T instruction manual Other Performance Problems, Excessive Noise, Slow Response