Model 9110T NOx Analyzer

EPA Protocol Calibration

10.6.1. SET UP FOR GPT MULTIPOINT CALIBRATION OF THE 9110T

The following 9110T features must be activated or set before calibration.

Calibrate the NO2 NO converter (see Section 9.1.4).

Set the reporting ranges for Independent mode (see Section 5.4.3.2).

Turn ON the automatic temperature/pressure compensation (TPC) feature (see Section 5.8).

Set the units of measure to ppb (see Section 5.4.3.4).

Figure 10-1: GPT Calibration System