Model 9110T NOx Analyzer

Instrument Maintenance

Figure 11-3: Ozone Cleanser Assembly

3.Remove the integrated screws with a Phillips screw driver and remove the scrubber manifold from the chassis.

4.Using a 9/16” wrench, remove both fittings from the cartridge.

5.Discard the glass wool.

6.Pour the contents of the scrubber manifold onto a sheet of white paper. If necessary, remove the plug to ensure that all the contents are poured out.

Notice any discoloration of the contents, which is usually white and slightly transparent.

The amount of discolored chemical (usually with yellow tint) may give you an indication of the lifetime of the chemical in your application.