Model 9110T NOx Analyzer

Calibration Procedures

SAMPLE

RANGE=500.0 PPB

NO=XXXX

<TST

TST>

CAL

 

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

 

 

CFG

DAS

RNGE PASS

CLK MORE

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

ENTER PASSWORD:818

 

 

 

8

1

 

 

8

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

0) DAS_HOLD_OFF=15.0 Minutes

<PREV NEXT> JUMP

EDIT PRNT EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Continue pressing NEXT until ...

SETUP X.X

09)CAL_ON_NO2=OFF

 

<PREV NEXT> JUMP

EDIT

PRNT EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODE FLD

CAL_ON_NO2=OFF

CONC

ON

 

 

 

ENTR EXIT

 

 

 

 

 

Use this button to

turn this feature

ON/OFF.

Press EXIT

3 times.