Model 9110T NOx Analyzer

 

Principles of Operation

 

 

 

 

Table 13-4: Relay PCA Status LED’s

 

 

 

 

 

 

 

 

 

 

 

 

LED

Color

Function

Status When Lit

Status When Unlit

 

 

 

(Energized State)

(Default State)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

D1

Red

Watchdog Circuit

Cycles ON/OFF every 3 Seconds

 

 

 

under direct control of the analyzer’s CPU.

 

 

 

 

 

 

 

 

 

D2

Yellow

Reaction Cell Heater

Heating

Not Heating

 

 

 

 

 

 

 

 

 

 

 

D3

Yellow

NO2 NO Converter Heater

Heating

Not Heating

 

 

 

 

 

 

 

 

 

 

 

D4

 

 

SPARE

 

 

 

 

D51

Yellow

Internal Span Gas Generator

Heating

Not Heating

 

 

 

 

 

Perm Tube Oven Heater

 

 

 

 

 

D6

 

 

SPARE

 

 

 

 

D7

Green

Zero/Span Valve

Valve OPEN to span gas flow

Valve OPEN to zero gas flow

 

 

 

D8

Green

Sample/Cal Valve

Valve OPEN to

Valve OPEN to

 

 

 

calibration gas flow

sample gas flow

 

 

 

 

 

 

 

 

 

D9

Green

Auto Zero Valve

Sample gas flow BYPASSES

Sample gas flow is routed

 

 

 

the reaction cell

THROUGH the reaction cell

 

 

 

 

 

 

 

 

 

D10

Green

NO/NOx Valve

Gas flow routed THROUGH

Gas Flow BYPASSES

 

 

 

NO2 NO converter

NO2 NO converter

 

 

 

 

 

 

 

 

 

D112

Green

Dual Span Gas

Valve OPEN to SPAN 1

Valve OPEN to SPAN2 inlet

 

 

 

 

 

Select Valve

gas inlet

 

 

 

 

D123

Green

Pressurized Span

Span gas flow SHUTOFF

Span gas flow OPEN

 

 

 

 

 

Shutoff Valve

 

 

 

 

 

D134

Green

Pressurized Zero

Zero gas flow SHUTOFF

Zero gas flow OPEN

 

 

 

 

 

Shutoff Valve

 

 

 

 

D14 - 16

 

 

SPARE

 

 

 

1

Only active when the optional internal span gas generator is installed.

 

 

 

2

Only active when the dual pressurized span option is installed.

 

 

 

3

Only active when one of the pressurized span gas options is installed.

 

 

 

4

Only active when one the pressurized zero gas option is installed.

 

 

 

 

 

 

 

 

 

 