Model 9110T NOx Analyzer

Setup Menu

a span gas with the concentration of 100 ppm can be used if the span gas passes through the same dilution steps as the sample gas, or;

a 5 ppm span gas must be used if the span gas IS NOT routed through the dilution system.

5.5.SETUP PASS: PASSWORD PROTECTION

The 9110T provides password protection of the calibration and setup functions to prevent unauthorized adjustments. When the passwords have been enabled in the PASS menu item, the system will prompt the user for a password anytime a password-protected function (e.g., SETUP) is selected. This allows normal operation of the instrument, but requires the password (101) to access to the menus under SETUP. When PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup (SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is enabled or disabled, a password (default 818) is required to enter the VARS or DIAG menus in the SETUP>MORE menu.

There are three levels of password protection, which correspond to operator, maintenance and configuration functions. Each level allows access to all of the functions in the previous level.

Table 5-2: Password Levels

PASSWORD

LEVEL

Null (000)

Operation

101

Configuration/Maintenance

 

 

818

Configuration/Maintenance

To enable passwords, press:

MENU ACCESS ALLOWED

All functions of the MAIN menu: TEST, GEN, initiate SEQ , MSG, CLR

Access to Primary Setup and Secondary SETUP Menus when PASSWORD is enabled.

Access to Secondary SETUP Submenus VARS and DIAG whether PASSWORD is enabled or disabled.

Teledyne Analytical Instruments

106

Page 106