Model 9110T NOx Analyzer

Data Acquisition System (DAS) and APICOM

7.1.3.9. The Compact Report Feature

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each parameter in one channel on a separate line, up to five parameters are reported in one line.

The COMPACT DATA REPORT generally cannot be accessed from the standard DAS front panel menu, but is available via the instrument’s communication ports by using APICOM or the analyzer’s standard serial data protocol.

7.1.3.10. The Starting Date Feature

This option allows the user to specify a starting date for any given channel when the user wants to start data acquisition only after a certain time and date. If the STARTING DATE is in the past (the default condition), the DAS ignores this setting and begins recording data as defined by the REPORT PERIOD setting.

The STARTING DATE generally cannot be accessed from the standard DAS front panel menu, but is available via the instrument’s communication ports by using APICOM or the analyzer’s standard serial data protocol.

7.1.3.11. Disabling/Enabling Data Channels

Data channels can be temporarily disabled, which can reduce the read/write wear on the Disk-on-Module (DOM).

To disable a data channel, go to the DAS>EDIT menu as shown in Section 7.1.3 then continue as follows:

Teledyne Analytical Instruments

180

Page 180