Model 9110T NOx Analyzer

Instrument Maintenance

CAUTION – QUALIFIED PERSONNEL ONLY

This procedure is carried out with the analyzer running and its cover removed.

1.Scroll the front panel display to show then test function to PMT.

2.Supply zero gas to the analyzer.

3.With the instrument still running, carefully remove the analyzer cover.

WARNING – ELECTRICAL SHOCK HAZARD

Do NOT touch any of the inside wiring with the metal cover or with your body.

Do NOT drop screws or tools into a running analyzer.

4.Shine a powerful flashlight or portable incandescent light at the inlet and outlet fitting and at all of the joints of the reaction cell as well as around the PMT housing.

The PMT value should not respond to the light, the PMT signal should remain steady within its usual noise floor.

5.If there is a PMT response to the external light, symmetrically tighten the reaction cell mounting screws or replace the 1/4” vacuum tubing with new, black PTFE tubing (this tubing will fade with time and become transparent).

Note

Often, light leaks are also caused by O-rings being left out of the

 

assembly.

6.If, during this procedure, the black PMT housing end plate for the Sensor Assembly is removed, ensure to replace the 5 desiccant bags inside the housing.

7.Carefully replace the analyzer cover. If tubing was changed, carry out a pneumatic leak check.

11.3.12.CHECKING FOR PNEUMATIC LEAKS

CAUTION - TECHNICAL INFORMATION

Do not exceed 15 psi when pressurizing the system during either Simple or

Detailed checks.

11.3.12.1. Simple Vacuum Leak and Pump Check

Leaks are the most common cause of analyzer malfunction. This section presents a simple leak check, whereas the next section details a more thorough procedure. The method described here is easy, fast and detects, but does not locate, most leaks. It also verifies the sample pump condition.

1.Turn the analyzer ON, and allow at least 30 minutes for flows to stabilize.

2.Cap the sample inlet port (cap must be wrench-tight).

3.After several minutes, when the pressures have stabilized, note the SAMP (sample pressure) and the RCEL (vacuum pressure) readings.

Teledyne Analytical Instruments

258

Page 258
Image 258