Model 9110T NOx Analyzer

Calibration Procedures

If the analyzer’s reporting range is set for the AUTO range mode, a step will appear for selecting which range is to be calibrated (LOW or HIGH). Each of these two ranges MUST be calibrated separately.

9.2.3.1. Setting the Expected Span Gas Concentration

Note

The expected concentrations for both NOx and NO are usually set to the

 

same value unless the conversion efficiency is not equal to 1.000 or not

 

entered properly in the conversion efficiency setting.

 

When setting expected concentration values, consider impurities in your

 

span gas source (e.g. NO often contains 1-3% NO2 and vice versa).

The NO and NOx span gas concentrations should be 80% of range of concentration values likely to be encountered in your application. The default factory reporting range setting is 500 ppb and the default span gas concentration is 40.0 ppb.

To set the span gas concentration, press:

Teledyne Analytical Instruments

200

Page 200
Image 200
Teledyne 9110T Setting the Expected Span Gas Concentration, Entered properly in the conversion efficiency setting