Model 9110T NOx Analyzer

Setup Menu

5.9.1. SIGNAL I/O

The signal I/O diagnostic mode allows a user to review and change the digital and analog input/output functions of the analyzer. Refer to Appendix A for a complete list of the parameters available for review under this menu.

IMPORTANT

IMPACT ON READINGS OR DATA

 

Any changes of signal I/O settings will remain in effect only until the

 

signal I/O menu is exited. Exceptions are the ozone generator override

 

and the flow sensor calibration, which remain as entered when exiting.

Access the signal I/O test mode from the DIAG Menu (Figure 5-4), then press:

DIAG

SIGNAL I / O

PREV NEXT JUMP

ENTR EXIT

 

 

Press NEXT & PREV to move between signal types.

DIAG I / O

0) EXT_ZERO_CAL=OFF

 

PREV NEXT JUMP

PRNT EXIT

 

 

 

EXAMPLE

 

 

 

 

 

DIAG I / O

JUMP TO: 12

 

1 2

 

ENTR EXIT

 

 

 

Press JUMP to go directly to a specific

signal

See Appendix A-4 for

a complete list of

available SIGNALS

EXAMPLE:

Enter 12 to Jump to

12) ST_SYSTEM_OK=ON

DIAG I / O

12) ST_SYSTEM_OK = ON

 

Exit to return

 

 

 

 

to the

PREV NEXT JUMP

ON

PRNT EXIT

DIAG menu

 

 

 

 

 

 

 

Toggle ON/(OFF) button to

 

Pressing the PRNT button will send a formatted

 

 

printout to the serial port and can be captured

 

change status.

 

with a computer or other output device.

 

 

 