Teledyne 9110T instruction manual Status Outputs, To test the status output electronics, Ozoneflow

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 287
Image 287

Model 9110T NOx Analyzer

Troubleshooting & Service

1.Use the Signal I/O function (see Section 12.1.3 and Appendix A) to view the value of REF_4096_MV and REF_GND.

If both are within 3 mV of nominal (4096 and 0), and are stable, ±0.2 mV then the basic A/D is functioning properly. If not then the motherboard is bad.

2.Choose a parameter in the Signal I/O function list (see Section 12.1.3) such as

OZONE_FLOW .

Compare this voltages at its origin (see the interconnect drawing and interconnect list in Appendix D) with the voltage displayed through the signal I/O function.

If the wiring is intact but there is a large difference between the measured and displayed voltage (±10 mV) then the motherboard is bad.

12.7.6.3.Status Outputs

 

V

+DC

Gnd

Figure 12-5: Typical Set Up of Status Output Test

To test the status output electronics:

1.Connect a jumper between the “D" pin and the “” pin on the status output connector.

2.Connect a 1000 ohm resistor between the “+” pin and the pin for the status output that is being tested.

3.Connect a voltmeter between the “” pin and the pin of the output being tested.

4.Under the DIAGSignal I/O menu (see Section 12.1.3), scroll through the inputs and outputs until you get to the output in question.

Teledyne Analytical Instruments

287

Page 287
Image 287
Teledyne 9110T instruction manual Status Outputs, To test the status output electronics, Ozoneflow