Teledyne 9110T instruction manual Manual Calibration Checks with Valve Options Installed

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 210
Image 210

Model 9110T NOx Analyzer

Calibration Procedures

9.4.2.MANUAL CALIBRATION CHECKS WITH VALVE OPTIONS INSTALLED

Analyzer display continues to cycle through all of the available gas measurements throughout this procedure.

SAMPLE

 

 

 

 

 

 

<TST

 

CAL

CALZ CALS

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Toggle

 

 

button until ...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

 

 

 

 

 

<TST

TST> CAL

 

 

CALS

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Record NOX, NO or NO2 zero point readings

Set the Display to show

the test

function.

This function calculates

the stability of the NO/NOx

measurement.

Wait until

falls below 0.5 PPB.

This may take several

minutes.

 

 

 

SAMPLE

 

 

 

 

 

 

 

<TST TST> CAL CALZ

SETUP

 

 

 

 

 

 

 

 

 

 

The

and/or

Wait until

 

 

 

 

buttons will appear at various

falls below 0.5 PPB.

 

 

 

 

 

 

This may take several

 

 

 

 

 

 

 

points of this process.

 

 

 

 

 

It is not necessary to press

 

 

 

 

 

 

 

 

minutes.

 

 

 

 

 

them.

 

 

 

 

 

 

 

 

 

 

 

 

Record NOX, NO, NO2 span point readings

 

 

 

 

 

 

 

 

 

 

 

Teledyne Analytical Instruments

210

Page 210
Image 210
Teledyne 9110T instruction manual Manual Calibration Checks with Valve Options Installed, Sample TST TST CAL Cals Setup