Teledyne 9110T Connections and Setup, Electrical Connections, Connecting Power, RS-485

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 41
Image 41

Model 9110TH NOx Analyzer

Getting Started

3.3. CONNECTIONS AND SETUP

This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2) connections for setup and preparing for instrument operation.

3.3.1. ELECTRICAL CONNECTIONS

Note

To maintain compliance with EMC standards, it is required that the cable

 

length be no greater than 3 meters for all I/O connections, which include

 

Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232,

 

and RS-485.

This section presents the electrical connections for AC power and communications.

3.3.1.1. Connecting Power

Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 A current at your AC voltage and that it is equipped with a functioning earth ground.

WARNING

ELECTRICAL SHOCK HAZARD

High Voltages are present inside the analyzers case.

Power connection must have functioning ground connection. Do not defeat the ground wire on power plug.

Turn off analyzer power before disconnecting or connecting electrical subassemblies.

Do not operate with cover off.

CAUTION

GENERAL SAFETY HAZARD

The 9110T analyzer can be configured for both 100-130 V and 210-240 V at either 47 or 63 Hz.

To avoid damage to your analyzer, ensure that the AC power voltage matches the voltage indicated on the analyzer’s model/specs label (Figure 3-4) before plugging the 9110T into line power.

3.3.1.2. Connecting Analog Inputs (Option)

The Analog In connector is used for connecting external voltage signals from other instrumentation (such as meteorological instruments) and for logging these signals in the analyzer’s internal DAS. The input voltage range for each analog input is 0-10 VDC.

Teledyne Analytical Instruments

41

Page 41
Image 41
Teledyne 9110T Connections and Setup, Electrical Connections, Connecting Power, Connecting Analog Inputs Option, RS-485