Model 9110TH NOx Analyzer

Communications Setup and Operation

Table 6-6: Default Hessen Status Flag Assignments

STATUS FLAG NAME3

WARNING FLAGS

DEFAULT BIT ASSIGNMENT

SAMPLE FLOW WARNING

OZONE FLOW WARNING

RCEL PRESS WARN

BOX TEMP WARNING

RCELL TEMP WARNING

IZS TEMP WARNING1

PMT TEMP WARN

CONV TEMP WARNING

INVALID CONC

OPERATIONAL FLAGS

0001

0002

0004

0008

0010

0020

0040

0080

8000

In MANUAL Calibration Mode

In ZERO Calibration Mode

In SPAN Calibration Mode

In WARMUP Mode

UNITS OF MEASURE FLAGS

0200

0400

0800

1000

UGM

MGM

PPB

PPM

SPARE/UNUSED BITS

0000

2000

4000

6000

0100

UNASSIGNED FLAGS (0000)

MANIFOLD TEMPERATURE2

HVPS WARNING

OZONE GEN OFF

FRONT PANEL WARN

SYSTEM RESET

ANALOG CAL WARNING

RELAY BOARD WARNING

CANNOT DYN ZERO

REAR BOARD NOT DETECTED

CANNOT DYN SPAN

AUTOZERO WARNING

Instrument is in MP CAL mode

1Only applicable if the optional internal span gas generator is installed.

2Only applicable if the 9110T is equipped with an oxygenator option.

3It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of similar flags, such as all temperature warnings, under the same status bit.

Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the assigned flags is active.

Teledyne Analytical Instruments

159

Page 159
Image 159