Model 9110TH NOx Analyzer

Setup Menu

5.9.2. ANALOG OUTPUT (DIAG AOUT)

Analog Output is used to verify functionality and accuracy of the analog outputs. The test forces all analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% increments. This test is useful to verify the operation of the data logging/recording devices attached to the analyzer.

Section 12.7.6.1 presents instructions for use in troubleshooting and service.

5.9.3. ANALOG I/O CONFIGURATION (DIAG AIO)

The 9110T analyzer comes equipped with four analog outputs. The first three outputs (A1 A2, & A3) carry analog signals that represent the currently measured concentrating of NOx, NO and NO2 (see Section 5.4.2.1). The fourth output (A4) outputs a signal that can be set to represent the current value of one of several test functions (see Table 5-9).

The following table lists the analog I/O functions that are available in the 9110T analyzer.

Table 5-5: DIAG - Analog I/O Functions

SUB MENU

 

FUNCTION

MANUAL

 

SECTION

 

 

 

 

 

 

 

Initiates a calibration of the A1, A2, A3 and A4 analog output channels that

 

AOUT CALIBRATED

determines the slope and offset inherent in the circuitry of each output.

5.9.3.1

These values are stored and applied to the output signals by the CPU

 

automatically.

 

 

Sets the basic electronic configuration of the A1 output (NOx Concentration).

 

 

There are four options:

 

 

RANGE1: Selects the signal type (voltage or current loop) and level of the

 

CONC_OUT_1

 

output

 

REC OFS: Allows them input of a DC offset to let the user manually adjust

 

 

 

the output level

5.9.2

 

AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature

 

 

 

CALIBRATED: Performs the same calibration as AOUT CALIBRATED, but

 

 

 

on this one channel only.

 

CONC_OUT_2

Same as for CONC_OUT_1 but for analog channel A2 (NO Concentration)

 

 

 

 

 

CONC_OUT_3

Same as for CONC_OUT_1 but for analog channel A3 (NO2 Concentration)

 

 

 

 

 

TEST OUTPUT

Same as for CONC_OUT_1 but for analog channel A4 (TEST CHANNEL)

5.9.4

 

 

 

AIN CALIBRATED

Initiates a calibration of the A-to-D Converter circuit located on the Motherboard.

5.9.3.10

 

 

 

 

XIN1

 

 

 

.

For each of 8 external analog inputs channels, shows the gain, offset,

 

.

engineering units, and whether the channel is to show up as a Test

5.9.3.11

.

function.

 

XIN8

 

 

 

 

 

 

 

Teledyne Analytical Instruments

117

Page 117
Image 117
Teledyne 9110T instruction manual Analog Output Diag Aout, Analog I/O Configuration Diag AIO