Model 9110T NOx Analyzer

Calibration Procedures

For each mode, there are seven parameters that control operational details of the SEQUENCE. They are:

 

Table 9-3: AutoCal Attribute Setup Parameters

 

 

 

 

ATTRIBUTE

 

 

ACTION

TIMER

 

Turns on the Sequence timer.

ENABLED

 

 

 

 

STARTING DATE

Sequence will operate after Starting Date.

STARTING TIME

Time of day sequence will run.

DELTA DAYS

 

Number of days to skip between each Sequence execution.

 

If set to 7, for example, the AutoCal feature will be enabled once

 

 

 

every week on the same day.

 

 

Number of hours later each “Delta Days” Sequence is to be run.

 

 

If set to 0, the sequence will start at the same time each day. Delta

DELTA TIME

 

Time is added to Delta Days for the total time between cycles.

 

This parameter prevents the analyzer from being calibrated at the

 

 

 

 

 

same daytime of each calibration day and prevents a lack of data for

 

 

 

one particular daytime on the days of calibration

 

 

Number of minutes the sequence operates.

 

 

This parameter needs to be set such that there is enough time for the

DURATION

 

concentration signal to stabilize.

 

The STB parameter shows if the analyzer response is stable at the

 

 

 

 

end of the calibration.

 

 

This parameter is logged with calibration values in the DAS.

 

 

Enable to do a calibration – Disable to do a cal check only.

CALIBRATE

 

For analyzers with internal span gas generators installed and

 

 

functioning, when used in US EPA applications, this setting must be

 

 

 

 

 

 

set to OFF.

 

 

LOW calibrates the low range, HIGH calibrates the high range. Applies

RANGE TO CAL

only to auto and remote range modes; this property is not available in

 

 

single and independent range modes.

IMPORTANT

IMPACT ON READINGS OR DATA

 

For US EPA controlled/related applications:

 

For analyzers used in US EPA controlled applications that have internal

 

span gas generators option installed, the CALIBRATE attribute must

 

always be set to OFF

 

Calibration of instruments used in US EPA related applications should

 

only be performed using external sources of zero air and span gas with an

 

accuracy traceable to EPA or NIST standards and supplied through the

 

analyzer’s sample port.

Teledyne Analytical Instruments

214

Page 214
Image 214
Teledyne 9110T For US EPA controlled/related applications, Always be set to OFF, Analyzer’s sample port, Attribute Action