Model 9110T NOx Analyzer

Troubleshooting & Service

Recalculate the converter efficiency (see previous section)

Accuracy of NO2 source (GPT or gas tank standard).

NO2 gas standards are typically certified to only ±2% and often change in concentrations over time. You should get the standard re-certified every year.

If you use the GPT calibration, check the accuracy of the ozone source.

Age of the converter.

The NO2 converter has a limited operating life and may need to be replaced every ~3 years or when necessary (e.g., earlier if used with continuously high NO2 concentrations).

We estimate a lifetime of about 10000 ppm-hours (a cumulative product of the NO2 concentration times the exposure time to that concentration).

This lifetime heavily depends on many factors such as:

Absolute concentration (temporary or permanent poisoning of the converter is possible).

Sample flow rate and pressure inside the converter.

Converter temperature.

Duty cycle.

This lifetime is only an estimated reference and not a guaranteed lifetime.

In some cases with excessive sample moisture, the oxidized molybdenum metal chips inside the converter cartridge may bake together over time and restrict air flow through the converter, in which case it needs to be replaced.

To avoid this problem, we recommend the use of a sample gas conditioner (Section 3.3.2.6).

Section 11.3.8 describes how to replace the NO2 converter cartridge.

With no NO2 in the sample gas and a properly calibrated analyzer, the NO reading is negative, while the NO2 reading remains around zero.

The converter is destroying NO and needs to be replaced.

With no NO2 in the sample gas and a properly calibrated analyzer, the NOX reading is significantly higher than the actual (gas standard) NO concentration.

The converter is producing NO2 and needs to be replaced.

12.7.11. SIMPLIFIED GPT CALIBRATION

This section describes how to determine the NO2 NO converter’s efficiency using a GPT method where the actual concentration of ozone is not a factor in the accuracy of the calculation.

This procedure is based on the Code of Federal Regulations, Title 40, Chapter I, subchapter C, Part 50, Appendix F.

Teledyne Analytical Instruments

296

Page 296
Image 296
Teledyne 9110T instruction manual Simplified GPT Calibration