Model 9110TH NOx Analyzer

Communications Setup and Operation

Example MODBUS Poll window:

6.7.2. HESSEN

The Hessen protocol is a multidrop protocol, in which several remote instruments are connected via a common communications channel to a host computer. The remote instruments are regarded as slaves of the host computer. The remote instruments are unaware that they are connected to a multidrop bus and never initiate Hessen protocol messages. They only respond to commands from the host computer and only when they receive a command containing their own unique ID number.

The Hessen protocol is designed to accomplish two things: to obtain the status of remote instruments, including the concentrations of all the gases measured; and to place remote instruments into zero or span calibration or measure mode. TAI’s implementation supports both of these principal features.

The Hessen protocol is not well defined; therefore, while Teledyne-API’s application is completely compatible with the protocol itself, it may be different from implementations by other companies.

6.7.2.1. HESSEN COMM PORT CONFIGURATION

Hessen protocol requires the communication parameters of the 9110T’s COMM ports to be set differently than the standard configuration as shown in the table below.

Table 6-4: RS-232 Communication Parameters for Hessen Protocol

PARAMETER

STANDARD

HESSEN

Baud Rate

300 – 19200

1200

Data Bits

8

7

Stop Bits

1

2

Parity

None

Even

 

 

 

Duplex

Full

Half

To change the baud rate of the 9110T’s COMM ports, See Section 6.2.2.

To change the remaining COMM port parameters listed in the table above, see Section 6.2.1, Table 6-1.

Note

Ensure that the communication parameters of the host computer are also

 

properly set.

 

Also, the instrument software has a 200 ms latency period before it

 

responds to commands issued by the host computer. This latency should

Teledyne Analytical Instruments

151

Page 151
Image 151
Teledyne 9110T Properly set, RS-232 Communication Parameters for Hessen Protocol, Parameter Standard Hessen