HP 1GSa/s Digitizing Oscilloscope 16532A manual Voltage Measurement Accuracy Test

Models: 1GSa/s Digitizing Oscilloscope 16532A

1 90
Download 90 pages 21.04 Kb
Page 33
Image 33

HP 16532A - Performance Tests

Procedure:

1.Using a BNC adapter, connect one end of the cable to the power supply. Connect the BNC tee, the blocking capacitor, and the shorting endcap to the other end of the cable. Refer to figure 3-1.

Figure 3-1. Voltage Measurement Accuracy Test

2.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given.

 

Menu

 

Selection

Setting

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Channel

 

Input

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

 

 

Probe

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1:1

 

 

Coupling

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 MΩ / DC

 

 

s/Div

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

500 ns

Trigger

 

Mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Immediate

Display

 

Mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Average

 

 

Average #

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8

 

 

Grid

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On

 

 

Markers

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set the Channel Coupling field to 1 MΩ/DC or damage to the equipment will result.

3-7

Page 33
Image 33
HP 1GSa/s Digitizing Oscilloscope 16532A manual Voltage Measurement Accuracy Test