HP 1GSa/s Digitizing Oscilloscope 16532A manual Offset Accuracy, Procedure Zero-Input Offset

Models: 1GSa/s Digitizing Oscilloscope 16532A

1 90
Download 90 pages 21.04 Kb
Page 35
Image 35

HP 16532A - Performance Tests

3-10. OFFSET ACCURACY

This test verifies the offset accuracy.

Specification:

± (1.0% of channel offset + 2.0% of full scale)

Equipment Required:

The following equipment is required for the test. Procedures are based on the model or part numbers rec- ommended.

 

 

 

Recommended

Equipment Required

 

Critical Specification

Model/Part Number

 

 

 

DC Power Supply

35.000 to + 35.000 Vdc, ± 1 mV resolution

HP 6114A

Digital Multimeter

Better than 0.1% accuracy

HP 3458A

Cable

50

Ω BNC (m to m) 48-inch

HP 10503A

Adapter

50

Ω BNC (f) to Dual Banana Plug

HP 1251-2277

 

 

 

 

 

 

 

 

Procedure:

Zero-Input Offset

1.Disconnect all cables and probes from the oscilloscope inputs. In the Waveform Selection menu de- lete channel 2. If channel 1 is not inserted, insert channel 1. Then set up the oscilloscope module according to the following parameters in the order given:

 

Menu

 

Selection

Setting

 

 

 

 

 

Channel

 

Input

 

1

 

 

Probe

 

1:1

 

 

v/Div

 

4 V

 

 

Offset

 

0 V

 

 

Coupling

 

1 MΩ / DC

 

 

s/Div

 

500 ns

Trigger

 

Mode

 

Immediate

Display

 

Mode

 

Average

 

 

Average #

 

32

 

 

Grid

 

On

 

 

Markers

 

On

 

 

 

 

 

 

 

 

 

 

2.With the DC power supply disconnected from the channel input, make sure the markers are near the center of the display.

3-9

Page 35
Image 35
HP 1GSa/s Digitizing Oscilloscope 16532A manual Offset Accuracy, Procedure Zero-Input Offset