HP 16532A - Performance Tests

3-6. DC CAL OUTPUT Test

The DC CAL OUTPUT port on the rear panel is used for instrument calibration and probe calibration. Though calibrator accuracy is not specified in the performance specifications, it must be within limits in or- der to provide accurate calibration.

Test Limits:

5.000 V ± 10 mV

Equipment Required:

The following equipment is required for the test.

 

 

 

Recommended

Equipment Required

 

Critical Specification

Model/Part Number

 

 

 

Digital Multimeter

0.1 mV resolution, better than 0.005% accuracy

HP 3458A

Cable

50

Ω BNC (m to m) 48-inch

HP 10503A

Adapter

50

Ω BNC (f) to Dual Banana Plug

HP 1251-2277

 

 

 

 

 

 

 

 

Procedure:

1.Use the BNC-to-banana adapter to connect the BNC cable between the multimeter and the oscillo- scope module DC CAL OUTPUT connector.

2.On the oscilloscope module set the following parameters in the order given:

 

Menu

 

Selection

Setting

 

 

 

 

 

Calibration

 

Mode

 

Service Calibration

 

 

Procedure

 

DC Cal BNC

 

 

DC Volts

 

0 V

 

 

 

 

 

 

 

 

 

 

3.The DVM should read close to 0.0000 V. Record the reading to four decimal places.

V1 = __________ .

4.In the Calibration menu set the DC Volts to 5 V.

5.The DVM should read close to 5.0000 V. Record the reading to four decimal places.

V2 = __________ .

6.In the Calibration menu set the DC Volts to 0 V.

7.Subtract V1 from V2. The difference should be between 4.990 and 5.010 V. Record the reading in the performance test record.