HP 1GSa/s Digitizing Oscilloscope 16532A manual Voltage Measurement Accuracy

Models: 1GSa/s Digitizing Oscilloscope 16532A

1 90
Download 90 pages 21.04 Kb
Page 32
Image 32

HP 16532A - Performance Tests

3-9. VOLTAGE MEASUREMENT ACCURACY

This test verifies the DC voltage measurement accuracy of the instrument. This procedure uses a dual cur- sor measurement that nullifies offset error, resulting in the following measurement:

± [(1.5% of full scale) + (0.016 x V/div)].

Specification:

±[(1.5% of full scale + offset accuracy) + (0.008 x V/div)]

Equipment Required:

The following equipment is required for this test. Procedures are based on the model or part numbers rec- ommended.

 

 

 

Recommended

Equipment Required

 

Critical Specification

Model/Part Number

 

 

 

DC Power Supply

-14 Vdc to + 14 Vdc, 0.1 mV resolution

HP 6114A

Digital Multimeter

Better than 0.1% accuracy

HP 3458A

Cable

50

Ω BNC (m to m) 48-inch

HP 10503A

Adapter

50

Ω BNC (f) to Dual Banana Plug

HP 1251-2277

(cable to power supply)

 

 

 

Adapter

50

Ω BNC tee (m)(f)(f)

HP 1250-0781

Blocking Capacitor

50 Ω BNC (m to f) 0.18 μF, 200 V

HP 10240B

 

 

 

 

 

 

 

 